COBAS 6000 C501 MODULE 05036453001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2010-02-12 for COBAS 6000 C501 MODULE 05036453001 manufactured by Roche Diagnostics.

Event Text Entries

[1426090] The user stated their results did not compare well with the peer group results for a therapeutic drug linearity survey. Of the data provided, the carbamazepine results for three samples, the vancomycin results for five samples and the valproic acid results for four sample were discrepant. All results are in ug/ml. The carbamazepine results were: sample 1 : 7. 7 and 7. 8 for a mean of 7. 75, the peer mean was 9. 37. Sample 2: 11. 0 and 10. 9 for a mean of 10. 95, the peer was 12. 82. Sample 3: 14. 0 and 13. 7 for a mean of 13. 85, the peer mean was 15. 96. The vancomycin results were: sample 1: results were 4. 3 and 4. 6 for a mean of 4. 45, the peer mean was 3. 25. Sample 2: results were 50. 5 and 50. 6 for a mean of 50. 55, the peer mean was 41. 27. Sample 3: results were 72. 9 and 72. 7 for a mean of 72. 80, the peer mean was 60. 43. Sample 4: results were 92. 0 and 91. 7 for a mean of 91. 87, the peer mean was 76. 65. Sample 5: results were 117. 7 and 114. 3 for a mean of 116. 00, the peer mean was 90. 65. The valproic acid results were: sample 1: results were 73. 4 and 78. 0 for a mean of 75. 70, the peer mean was 70. 89. Sample 2: results were 99. 6 and 97. 2 for a mean of 98. 40, the peer mean was 98. 80. Sample 3: results were 120. 1 and 112. 0 for a mean of 116. 05, the peer mean was 125. 80. Sample 4: results were 141. 6 and 153. 8 for a mean of 147. 70, the peer mean was 147. 13. The samples were tested in microcups. The carbamazepine reagent lot number was 584903, the vancomycin reagent lot number was 14455000 and the valproic acid lot number was 14454900. The user retested the same samples on (b) (6) 2010 and (b) (6) 2010 with new lot numbers of reagents and results were within the acceptable range. The carbamazepine reagent lot number 587925. The vancomycin and valproic acid lot number were not provided. The carbamazepine repeat results were: sample 1 : 9. 714 and 9. 770. Sample 2: 13. 305 and 13. 363. Sample 3: 16. 381 and 17. 139. The vancomycin results were: sample 1: 4. 7 and 5. 1. Sample 2: 41. 6, 42. 3 and on (b) (6) 2010 was 40. 4. Sample 3: 61. 2, 61. 3 and on (b) (6) 2010 was 57. 0. Sample 4: 78. 2, 79. 9 and on (b) (6) 2010 was 71. 6. Sample 5: 96. 5 with a data flag, 100. 3 with a data flag and on (b) (6) 2010 was 86. 3. The valproic acid results were: sample 1: 89. 8, 92. 0 and on (b) (6) 2010 was 77. 0. Sample 2: 123. 0, 125. 0 and on (b) (6) 2010 was 103. 4. Sample 3: 164. 9 with a data flag, 171. 9 with a data flag and on (b) (6) 2010 was 120. 9. Sample 4: 196. 5 with a data flag, 171. 9 with a data flag and on (b) (6) 2010 was 143. 5. The user stated no patient samples were affected by the issue. The user stated he felt the cause may be a reagent issue or a poor calibration. The field application specialist calibrated the assays with a new calibrator and recovered qc values close to the mean. He then retested the linearity material and results recovered in range.
Patient Sequence No: 1, Text Type: D, B5


[8508821] It was unknown if the initial reporter sent report to the fda.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2010-00979
MDR Report Key1598505
Report Source05,06
Date Received2010-02-12
Date of Report2010-02-12
Date of Event2009-12-01
Date Facility Aware2010-01-19
Date Mfgr Received2010-01-19
Date Added to Maude2010-02-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA ERIC KOLODZIEJ
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175212834
Manufacturer G1HITACHI HIGH TECH. CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOBAS 6000 C501 MODULE
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeKLT
Date Received2010-02-12
Model NumberNA
Catalog Number05036453001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2010-02-12

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.