COBAS 6000 C501 MODULE 05860636001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-10-20 for COBAS 6000 C501 MODULE 05860636001 manufactured by Roche Diagnostics.

Event Text Entries

[90768595] (b)(4). (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[90768596] The customer complained of questionable trigl triglycerides (trigl) results for two patient samples. For patient 1 the initial trigl result was 580 mg/dl. The repeat result was 87 mg/dl. For patient 2 the initial trigl result was 600 mg/dl. The repeat result was 98 mg/dl. The initial results were reported outside of the laboratory. The repeat testing was performed on another analyzer. The repeat results were deemed to be correct. There was no allegation of an adverse event. The trigl reagent lot was 23611201 with an expiration date of 30-apr-2018. A specific root cause could not be identified. Further investigation excluded a general reagent problem due to calibration and qc results being acceptable. It was noted that the customer changes the calibration factors after performing quality control testing. The customer was informed that this not recommended. The alarm trace for the analyzer showed frequent probe alarms. The customer cleaned the probes and is having no further issues.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-02386
MDR Report Key6965215
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-10-20
Date of Report2017-10-20
Date of Event2017-10-05
Date Mfgr Received2017-10-06
Date Added to Maude2017-10-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOBAS 6000 C501 MODULE
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeJGY
Date Received2017-10-20
Model NumberC501
Catalog Number05860636001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-20

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.