MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2000-02-07 for SYNTHESIS 1715 0001715097 manufactured by Instrumentation Laboratory Co..
[146602]
A customer complaint was received that the il synthesis model 1715 reported an erroneous cohb result of 18% and that the instrument did not flag the error. The quality controls also reported within specification. A site visit by an il field service representative (fsr) showed that the instrument was not being maintained per labeling instructions and as a result a film formed in the cuvette window. Furthermore, cleaning pathway errors had been flagged. After cleaning by the fsr, instrument results returned to normal. Based on a follow-up conversation between the site and the fsr, it was concluded that no pt was treated based on the erroneous cohb result.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1217183-2000-00001 |
MDR Report Key | 261680 |
Report Source | 05 |
Date Received | 2000-02-07 |
Date of Report | 2000-02-07 |
Date of Event | 2000-01-06 |
Date Mfgr Received | 2000-01-07 |
Device Manufacturer Date | 1998-02-01 |
Date Added to Maude | 2000-02-11 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 0 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 0 |
Manufacturer Contact | CAROL MARBLE |
Manufacturer Street | 113 HARTWELL AVE. |
Manufacturer City | LEXINGTON MA 02421 |
Manufacturer Country | US |
Manufacturer Postal | 02421 |
Manufacturer Phone | 7818614467 |
Manufacturer G1 | * |
Manufacturer Street | * |
Manufacturer City | * |
Manufacturer Country | * |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | SYNTHESIS |
Generic Name | BLOOD GAS/ELECTROLYTE/GLUCOSE/CO-OXIMETER |
Product Code | JKS |
Date Received | 2000-02-07 |
Model Number | 1715 |
Catalog Number | 0001715097 |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Implant Flag | N |
Date Removed | A |
Device Sequence No | 1 |
Device Event Key | 253432 |
Manufacturer | INSTRUMENTATION LABORATORY CO. |
Manufacturer Address | VIALE DELL'INDUSTRIA 3 PADERNO, DUGNANO IT 20037 |
Baseline Brand Name | SYNTHESIS |
Baseline Generic Name | BLOOD GAS/ELECTROLYTE/GLUCOSE/CO-OXIMETER |
Baseline Model No | 1715 |
Baseline Catalog No | 0001715097 |
Baseline ID | NA |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2000-02-07 |