HEMOCHRON JR. MICROCOAGULATION ACT LOW RANGE JACT-LR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,health professional,use report with the FDA on 2015-06-24 for HEMOCHRON JR. MICROCOAGULATION ACT LOW RANGE JACT-LR manufactured by International Technidyne Corp..

Event Text Entries

[18040838] Healthcare professional reported inconsistent patient results with the hemochron signature elite and act low range system. A (b)(6) male was receiving heparin anticoagulation via a continuous infusion plus iv bolus doses during a cardiac ablation procedure for atrial fibrillation. Three hemochron signature elite instruments were run side by side with one reagent cuvette lot of act-lr in a correlation study. The target act was greater than 300 seconds. During the procedure, act readings wore discrepant on a few occasions between the three instruments; however two values differed by more than what would be expected. Electronic and liquid quality controls passed successfully on all three instruments. The procedure was completed successfully and no patient injury or adverse events were reported.
Patient Sequence No: 1, Text Type: D, B5


[18140364] (b)(4). The serial numbers of the hemochron signature elite instruments used during this procedure were (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[68476677] Mdr follow up #1 references itc complaint number (b)(4) and provides the results from testing reserved samples of the same reagent lot by lsr 2015-css-032. Results: precision outside of prespecified acceptance criteria at assay upper range confirms the complaint of inconsistent patient results.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2248721-2015-00053
MDR Report Key4870774
Report Source05,06,HEALTH PROFESSIONAL,USE
Date Received2015-06-24
Date of Report2015-06-01
Date of Event2015-06-01
Date Mfgr Received2015-08-20
Device Manufacturer Date2014-12-01
Date Added to Maude2015-06-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJON MCDERMED
Manufacturer Street8 OLSEN AVENUE
Manufacturer CityEDISON NJ 08820
Manufacturer CountryUS
Manufacturer Postal08820
Manufacturer Phone8582632490
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHEMOCHRON JR. MICROCOAGULATION ACT LOW RANGE
Generic NameTEST, TIME, ACTIVATED CLOTTING TIME
Product CodeJBP
Date Received2015-06-24
Model NumberJACT-LR
Catalog NumberJACT-LR
Lot NumberM4JLR201
Device Expiration Date2016-03-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINTERNATIONAL TECHNIDYNE CORP.
Manufacturer AddressEDISON NJ 08820 US 08820


Patients

Patient NumberTreatmentOutcomeDate
10 2015-06-24

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.