BE-HQV 33800#COMPLETE PACK BE-00175 701002503

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-09-23 for BE-HQV 33800#COMPLETE PACK BE-00175 701002503 manufactured by Maquet Cardiopulmonary Ag.

Event Text Entries

[55794700] (b)(4). Maquet cardiopulmonary ag is aware of similar complaints from this product. Similar products, showing a similar malfunction, have been tested. The tested products has the same leakage at the same position. This could be determined as bonding failure. Visual inspection, tightness test acc. To lv 203 and tightness test with roller pump have been performed. Results: during tightness test acc. To lv 203 the exact position of the leak in the arterial filter between the cover and the housing could be detected (improper ultrasonic welding). During testing with roller pump leakage at quart could also be detected but the exact position was not found during this kind of test. The existing liquid collected at the lowest point of the filter and drained. Most possible root cause could be the bad bonding between cover and filter body. The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process. Due to this no further investigation initiations will be completed at this time.
Patient Sequence No: 1, Text Type: N, H10


[55794701] It was reported that prime a leak occurred at by opening the bypass. No known consequences to the patient were reported. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010762-2016-00588
MDR Report Key5972617
Date Received2016-09-23
Date of Report2016-08-31
Date of Event2016-08-16
Date Mfgr Received2016-08-31
Device Manufacturer Date2016-05-01
Date Added to Maude2016-09-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer StreetMAQUET CARDIOPULMONARY AG KEHLER STRASSE 31
Manufacturer City76437 RASTATT
Manufacturer CountryGM
Manufacturer Phone4972229321
Manufacturer G1BERND RAKOW
Manufacturer StreetMAQUET CARDIOPULMONARY AG KEHLER STRASSE 31
Manufacturer City76437 RASTATT
Manufacturer CountryGM
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBE-HQV 33800#COMPLETE PACK
Generic NameFILTER, BLOOD, CARDIOPULMONARY BYPASS, ARTERIAL LINE
Product CodeDTM
Date Received2016-09-23
Model NumberBE-00175
Catalog Number701002503
Lot Number92189712
Device Expiration Date2018-05-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMAQUET CARDIOPULMONARY AG
Manufacturer AddressRASTATT GM


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-23