MAR COR COMPLETE WATER TREATMENT SYSTEM PVC LOOP

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2016-10-17 for MAR COR COMPLETE WATER TREATMENT SYSTEM PVC LOOP manufactured by Mar Cor Purification.

Event Text Entries

[57682069] Mar cor (mcp) technician arrived on a sunday to perform a dialysis loop disinfect service. Notified nurses on-site prior to starting the disinfection and received permission to start the monthly disinfect. After the loop disinfection and during the dialysis loop rinse out, the mcp technician was performing a disinfectant residual check of all the patient rooms on the 6th floor. During this rise out/residual check it was discovered a dialysis technician had started a patient on dialysis. The dialysis machine was disconnected from the water loop and the water loop port tested positive for disinfectant. Rinse out of the water loop port was performed until the residual disinfectant check showed negative. The clinic staff was notified immediately of this incident. Mcp has spoken with the clinic and has not been notified of any patient injury or patient requiring medical attention.
Patient Sequence No: 1, Text Type: N, H10


[57682070] Potential of exposure of one dialysis patient to minncare. No patients were hospitalized.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3019131-2016-00001
MDR Report Key6034695
Report SourceCOMPANY REPRESENTATIVE
Date Received2016-10-17
Date of Report2016-10-17
Date of Event2016-09-18
Date Mfgr Received2016-09-18
Date Added to Maude2016-10-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MEGAN DICKEY
Manufacturer Street14550 28TH AVENUE NORTH
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal55447
Manufacturer Phone7635533300
Manufacturer G1MAR COR PURIFICATION
Manufacturer Street14550 28TH AVENUE NORTH
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal Code55447
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMAR COR COMPLETE WATER TREATMENT SYSTEM
Generic NameSUBSYSTEM, WATER PURIFICATION
Product CodeFIP
Date Received2016-10-17
Model NumberPVC LOOP
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMAR COR PURIFICATION
Manufacturer Address14550 28TH AVENUE NORTH MINNEAPOLIS MN 55447 US 55447


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-10-17