ALLURA XPER FD10 722010

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-12-21 for ALLURA XPER FD10 722010 manufactured by Philips Healthcare.

Event Text Entries

[131484583] When the investigation has been completed philips will inform the fda.
Patient Sequence No: 1, Text Type: N, H10


[131484584] Philips has received through (b)(6) a report submitted ((b)(4)) by a customer. In this report the customer reported:? The x-ray tube in room 5 had been repaired, updated during the day of (b)(6) 2018. Was reported ready for use. We re-booted the system at the beginning of the night shift due to that the table was hard to operate (a creaking sound). Hereafter there was an acute patient on the table without x-ray problems. After some hours a new acute patient arrive. Before this patient was brought onto the table we discovered that the table was floating? Meaning that it was moving by the leased touch. Therefore it was not possible to complete the kag/pci on this very bad acute patient. Had to move the patient to another room which took approx. 10 min. Normally? When treating acute patients? It takes approx.. 5 min when the patient arrive at the room until the procedure can start?. Philips has initiated an investigation of this complaint.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003768277-2018-00098
MDR Report Key8188893
Date Received2018-12-21
Date of Report2018-12-06
Date Mfgr Received2018-12-06
Device Manufacturer Date2010-08-09
Date Added to Maude2018-12-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DUSTY LEPPERT
Manufacturer StreetVEENPLUIS 4-6 P.O. BOX 10.000
Manufacturer CityBEST 5680DA
Manufacturer CountryNL
Manufacturer Postal5680 DA
Manufacturer G1PHILIPS MEDICAL SYSTEMS
Manufacturer Street3000 MINUTEMAN ROAD
Manufacturer CityANDOVER MA 01810
Manufacturer CountryUS
Manufacturer Postal Code01810
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameALLURA XPER FD10
Generic NameSYSTEM, X-RAY, ANGIOGRAPHIC
Product CodeIZI
Date Received2018-12-21
Model Number722010
Catalog Number722010
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS HEALTHCARE
Manufacturer AddressVEENPLUIS 4-6 P.O. BOX 10.000 BEST 5680DA NL 5680 DA