IFUSE IMPLANT SYSTEM 7055M-90

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-01-13 for IFUSE IMPLANT SYSTEM 7055M-90 manufactured by Si-bone, Inc..

Event Text Entries

[174840627] Based on the information provided, review of the surgical technique manual, ifu, certificates of analysis and fmea, the most probable root cause is user error; improper implant size selection, using too long of an implant.
Patient Sequence No: 1, Text Type: N, H10


[174840628] The patient had left side si joint arthrodesis in (b)(6) 2019 where three implants were installed. The patient reported radicular leg pain following the procedure. The surgeon determined that the superior positioned implant was impinging on the neuroforamen. Later that same day, the surgeon performed a revision procedure where he removed the impinging implant and replaced it with a shorter implant of the same type. No other preexisting implants were adjusted or removed. The patient's radicular pain symptoms resolved following the revision procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007700286-2019-00162
MDR Report Key9581375
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-01-13
Date of Report2020-01-09
Date of Event2018-12-14
Date Mfgr Received2019-12-23
Device Manufacturer Date2017-07-12
Date Added to Maude2020-01-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactW. RECKLING, M.D.
Manufacturer Street471 EL CAMINO REAL SUITE 101
Manufacturer CitySANTA CLARA CA 950504482
Manufacturer CountryUS
Manufacturer Postal950504482
Manufacturer Phone4082070700
Manufacturer G1SI-BONE, INC.
Manufacturer Street471 EL CAMINO REAL SUITE 101
Manufacturer CitySANTA CLARA CA 950504482
Manufacturer CountryUS
Manufacturer Postal Code950504482
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIFUSE IMPLANT SYSTEM
Generic NameORTHOPEDIC ROD
Product CodeOUR
Date Received2020-01-13
Model Number7055M-90
Lot Number2591268
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSI-BONE, INC.
Manufacturer Address471 EL CAMINO REAL SUITE 101 SANTA CLARA CA 950504482 US 950504482


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-01-13